Welcome to CHINOLAXSONS - pioneer in instrumentation, controls & system.
CHINO Laxsons
 

Infrared Radiation Thermometers

On-Line Type

IR-A Series Digital Infrared Radiation Thermometers
The IR-A series is digital infrared radiation thermometers covering the wide range from -50 to 3500. The high-performance thermometers ensure highly precious and highly reliable measurements in the wide applications from iron/steel and ceramic fields to high technology fields of new materials and semiconductors.
Features
  • Standard functions conform to the systemized radiation thermometry.
  • Wide measuring range and optional scaling of scale range
  • Interchangeable detectors and converters
  • Compact and lightweight
  • Sharp moving focus type optical system
  • Highly reliable on-line sensor
Models
[Thermometer]
  • Radiation thermometer for low temperature

Types

Distance factors

Models

Standard

50

IR-AL0CB

For small diameter with short distance

2/100mm

IR-AL8CB-1

4/200mm

IR-AL8CB-2

  • Radiation thermometer for medium/high temperature

Distance factors

Detectors

PbSe

PbS

Ge

Si

50

--

--

IR-AP0CG

IR-AP0CS

100

IR-AP1CE

IR-AP1CP

IR-AP1CG

IR-AP1CS

200

--

--

--

IR-AP2CS

300

--

IR-AP3CP

IR-AP3CG

IR-AP3CS

  • 2-color thermometer

Distance factors

Detectors

Ge

Si

35

IR-AQ8CG

--

50

IR-AQ0CG

--

100

IR-AQ1CG

IR-AQ1CS

200

--

IR-AQ2CS

300

--

IR-AQ3CS

[Converter]

Types

Models

Panel-back-mount type

IR-GAB

Panel mount type

IR-GAP

Wall mount type

IR-GAW

Panel mount type with multi-function

IR-GAG

Specifications
[Thermometer]

Items

For low temperature

IR-AL

For medium/high temperature
IR-AP

2-color type

IR-AQ

Measuring systems

Wide band radiation thermometer

Narrow band radiation thermometer

2-color thermometer

Temperature ranges

-50 to 1000

400 to 1300 and other ranges in 100 to 3500

800 to 1800 and other ranges in 400 to 3100

Detectors

Pyroelectric

PbSe, PbS, Ge, Si

Ge, Si

Spectral ranges

8 to 13m

PbSe: 4m

PbS: 2m

Ge: 1.6m

Si: 0.96m

Ge: 1.5/1.65m
Si: 0.8/0.97
m

Accuracy ratings

Less than 200: 2

More than 200: 2% of reading

PbSe, PbS, Ge:

Less than 500: 3
500 to 1000: 5
More than 1000: 0.5% of reading
Si:
Less than 1500: 0.5% of reading
1500 to 2000: 1% of reading
More than 2000: 2% of reading

400 to 700: 1% of reading
700 to 1500: 0.5% of reading
1500 to 2000: 1% of reading
More than 2000: 2% of reading

Resolutions

1 (for the measured temperature more than 50)

Ge, Si: 1.5 (less than 400), 1 (more than 400)
PbSe, PbS: 2 (less than 300), 1 (more than 300)

2

Repeatability

Within 1

Response time

0.5 second (95% response)
(on output of converter)

0.2 second (95% response)
(on output of converter)

0.3 second (95% response)
(on output of converter)

Distance factors

50

50, 100, 200, 300

35, 50, 100, 200, 300

Measuring distance

0.5 m to

Measuring diameter

Measuring distance/Distance factor

Collimation

Direct viewfinder

Operating temperature

0 to 50
0 to 150 (on water-cooling by water cooling jacket)

Connection

Terminal connection or connector connection

Connection cable length

Up to 200m by exclusive cable

Weight

About 1.9kg

[Converter]

Display

Panel-back-mount type
IR-GAB

Panel mount type

Wall mount type
IR-GAP, IR-GAW

Panel mount type with multi-function
IR-GAG

Emissivity (Emissivity ratio)

0.100 to 1.000 (0.800 to 1.200)

Modulator

Delay: Trace of average value (smoothing)

Real: Original signal
Peak: Trace of maximum value
Delay: Trace of average value (smoothing)
Valley: Trace of minimum value

Display

--

LED 4 digits (red)

Output signal

4 to 20mADC (Load resistance: less than 550)
0 to 10mVDC (Output resistance 10
)
Isolate output 

4 to 20mADC (Load resistance: less than 550)
0 to 10mVDC (Output resistance 10
)
Independent isolate output 

Output scaling

By standard range

Zero and span are freely adjustable 1 step.

Output renewal cycle

0.1 second

Alarm output

--

High/low independent setting
Output terminals: 2 sets of H, C, and L terminals
Contact capacity: 100VAC, 1A (resistive load)

Power supply

100, 110, 200 or 220VAC (to be specified), 50/60Hz

Power consumption

Maximum 30VA

Maximum 40VA

Operating temperature range

0 to 50

Calculation

--

--

Block processing, reflection correction, emissivity (ratio) program, 2-input selection, emissivity calibration

Weight

About 2kg

Panel mount type:
About 2.5kg
Wall mount type:

About 6kg

About 4kg

Option

--

--

RS-232C or RS-422A
(to be specified)